This tritium sampler brings original solutions in the exploitation of trapping systems for tritium gas and tritiated water. This collection system is mainly used for the detection of chimney rejects and TRITIUM wastes degassing. The realization of this machine has been essentially focused on optimum sampling efficiency associated with ergonomics adapted to all utilization conditions. The technician can come less frequently to collect the liquid solution (monthly instead of a usual weekly frequency) thanks to a cooling system water circuit that will prevent all water loss by evaporation in the feeding-bottles. Its original tightness dynamics system on the two feeding-bottles, authorizes a manipulation (on/ off) particularly confoftable of this device in all circumstances without pressing back nor loss of solution. A microprocessor pilots and manages all functioning parameters, insuring thus a total security in its utilization during long periods.
Working principle
This device traps the tritiated water vapor by a principle of paddling in two feeding-bottles. The collector sucks up atmospheric air through a meter which precisely measures the quantity of aspired air. This air runs successively through two bottles that trap the tritiated water vapor. After a certain time, the tritiated water contained in the bottles is measured in a laboratory. The quantity of tritium measured is related to the volume air which passed in the equipment.
Quality norm
IN COMPLIANCE WITH THE NF M60-312 NORM
Key Advantages
1. GOODTRAPPING EFFICIENCY.
2. COOLING SYSTEM TO INCREASE SAMPLING LENGTH (option).
3. GOOD PRICE.
4. CONSTANT EVOLUTION OF THE PRODUCT.
5. VERY ROBUST.
6. EASY TO USE.
7. CONNECTABLE TO ALL SAMPLING LINES.
ELECTRICAL SPECIFICATIONS
Voltage requirement : 220 Volts. (±10%).
Power consumption : 300 Watts (700 Watts, with cooling system option).
Frequency voltage: 50 Hz (±10%).
Running fully piloted by microprocessor : Airflow (10 to 55 liters/hour) and detection of tightness defect of bottles.
Heap of gases posted by the machine ( until 25,000 m3).
Cooling system : +5°C to +15°C (depending on the ambient temperature).
Circulation of the cold liquid : automatic error detection.
Circulation pump of the cold liquid (functioning).
Alarm Buzzer for all defects detected.
Electrical output by : RS 232. (J.BUS system optional).
Electrical signal TOR card - alarm state relay (option).
Detection and identification of defects at a distance (via RS 232).
Differential circuit-breaker (cut current = 30mA).
Display : to crystals liquidate type matrical 2 lines of 16 characters.
Simultaneous display of: Oven and cold temperature. Instantaneous airflow and volume accrued. Duration of the trapping and alarm.
Resumption of running witnesses of the cold group by LED diode on front panel of MARC 5000 equipment.
Power consumption by the cold system: 400 Watts.
MECHANICAL SPECIFICATIONS
Monoshell BOX in alloy of aluminium.
Front panel in LEXAN anti-stripe material.
Box painted : special powder paint, seed look.
Main materials constituting the mechanical parts are; PVC, POLYACETAL, POLYAMID, NYLON, Alloy of Aluminium, STAINLESS Steel, brass nickel-plated.
Feeding-bottles : Made of glass. Content is about 250 ml (the admissible height of the bottle is between 158 and 163mm).
Pump of air aspiration : 10 to 55 liters per hour.
Electronic airflow sensor (accurate= 1% calibrated).
Meadow-paper filter to the admission of air (= 45 mm).
Operating temperature : +0°C to + 45°C.
Storage temperature : -25°C to + 70°C. Running the equipment below 0°C is possible as long as diluted alcohol (ethanol) is added to the water in the feeding bottles.
Weight of the machine : without cold system option:18Kg. With cold system option : 29 Kg.
Carriage : 3 handles system for a confortable transportation by 2 people.
Dimensions : Length= 580 mm, Height= 265 mm, Depth= 210 mm
[**] Cooling system of water (optional) : Pump of water circulation (2 liters per minute).
Cold system with condenser. Thermal dissipation insured by a turbine to air. Mini temperature obtained = +2°C to +30°C under ambient temperature or +9°C to +45°C under ambient temperature.